Today's Tip

Paper Thermometer Reveals Temperature Spikes

Each tip published will earn the sender $50. Submit your tip


Book Bits

Pressure Indicators Provide Early Warning

From "Total Contamination Control '98"

Differential pressure indicators are used to give warning of impending filter blockage. The pressure at which actuation occurs depends upon the setting of the bypass valve chosen and is usually set at about 70 percent of the bypass cracking pressure. This gives a safety margin so that element change can be effected at a convenient time before bypassing occurs.

More information about the book "Total Contamination Control '98"


Q & A

Particle Counter Results Differ

"I have recently purchased an on-site portable particle counter allowing me to take immediate action on out-of-condition results. However, my instrument results seem to be lower than the results I receive from the laboratory. Is my new instrument at fault?"

Ownership of an on-site instrument is indeed a valuable addition to your program. However, do bear in mind the difference between taking a bottle sample and attaching your instrument directly online with the system.

Effectively, your instrument is giving you a direct insight to the system cleanliness without the corruption of potentially poor sampling procedures such as insufficient test port flushing, ambient contamination levels, and questionable bottle cleanliness.

My experience suggests that a difference of three ISO code values is the norm when comparing a laboratory result to an instrument, and may stretch to even wider differences. A difference of three codes is effectively a sample that is eight times dirtier - which when concentrated in a small sample bottle, is understandable.

However, proactive practitioners have successfully used laboratory bottle sampling to monitor their cleanliness levels, but I suggest you either make adjustments in your target cleanliness levels for the instrument, or set the targets based on bottle sampling, but monitor for sudden changes in the instrument cleanliness results that may not exceed the target caution/ critical levels.

Martin Williamson, Senior Technical Consultant, Noria UK Ltd.

Submit your question here.


Suggestions, Questions and Tip Submissions

Submit questions or tips.

Other correspondence:

Noria Corporation
1328 E. 43rd Court
Tulsa, OK 74105 USA
Phone: 918-749-1400
Fax: 918-746-0925

Noria publishes two magazines with complimentary subscriptions in the U.S., Europe and Canada.

Machinery Lubrication Magazine and Practicing Oil Analysis Magazine

Lube-Tips is also published in Spanish. Request your subscription at informes@noria.com

Copyright © 2003, Noria Corporation. All rights reserved. If you would like to reproduce a Lube-Tip on your Web site, you must use the entire issue (without sponsorship messages and the training calendar). All links must work. For an example of how you can include Lube-Tips content on your Web site, go to: http://www.lube-tips.com/example.asp. The presence of advertising in Lube-Tips does not constitute an endorsement of the products or services in such ads. Further, because results will vary widely based on a number of factors, Noria Corporation cannot warrant the results, the accuracy or the completeness of any material published herein.

Lube-Tips is published by Noria Corporation. Oil Analysis and Lubrication Experts